h. Cleaning System Data – Cleaning Cycle Reports (Reviewing Cycle Events)
Notes 1. The cycle events that occur during a cleaning cycle (such as OpCodes/steps, messages, operator actions, or alarms) can be reviewed in this tab. This is useful for troubleshooting and diagnosing issues that occur during a cleaning cycle such as seeing what operator actions were taken or what events occurred prior to an alarm or aborted cleaning cycle. The following data is presented: a. Description – the event name that occurs. These are classified into 4 categories and color coded per below: i. OpCodes (e.g., Fill Sump – Hot, Pre-Rinse – Heat) ii. Operator Actions (e.g., Start Button Pressed, Manual Mode) iii. Messages (e.g., Temperature Hold, Fill Hold) iv. Alarms (e.g., Low Supply Flow, Low Instrument Air Pressure) 2. This data can also be exported to a table – in the lower right select “Export Trend Data” to download a .csv file of the tabled data.