OPERATING REQUIREMENTS
UltraFlow 45
UltraFlow 110
Instrument Air
½” NPT, 10 scfm @ 90 psi
½” NPT, 10 scfm @ 90 psi
Two 1” tri-clamps, WFI, DI, potable ≤ 2 gpm @ 25 psi, 20°–80 °C
Two 1” tri-clamps, WFI, DI, potable ≤ 2 gpm @ 25 psi, 20°–80 °C
Water Supply
Drain
2” tri-clamp (controllable drain rate)
3” tri-clamp (controllable drain rate)
Dry Weight
900 lbs (approximate)
1,400 lbs (approximate)
Electrical Power (w/ Electric Heat) Electrical Power (w/ Optional Steam Heat)
12 kW, 27 amps (standard) or 24 kW, 43 amps (optional) @ 460V AC, 3PH
15 kW, 50 amps (standard) or 30 kW, 68 amps (optional) @ 460V AC, 3PH
11 amps @ 460V AC, 3PH
27 amps @ 460V AC, 3PH
Plant Steam
¾” flange, 195 lbs/hr @ 50 psi
1 ½” flange, 540 lbs/hr @ 50 psi
Plant Condensate
½” flange
1” flange
CIP Supply
1 ½” tri-clamp, 2–45 gpm @ 50 psi
2” tri-clamp, 5–110 gpm @ 60 psi
CIP Return
2” tri-clamp, 2–45 gpm @ 8.5’ of head @ 80 °C 3” tri-clamp, 5–110 gpm @ 11’ of head @ 80 °C
Vent/Overflow
2” tri-clamp
2” tri-clamp
SEPARATOR CHAMBER: SMALL SIZE, BIG PERFORMANCE
Water Flow Air Flow
The combination of air and CIP return solution enters the sidewall port of the separator where centrifugal action separates air (upwards) and solution (downwards) to maintain adequate supply conditions for the CIP supply centrifugal pump.
Air Separation Chamber
7