QUICK TIPS The ALVIM Biofilm Sensor is able to detect bacterial settlement since its first phases. The image below shows early biofilm settlement and the resulting signal indicator.

GOOD TO KNOW • Wetted Materials: 316Lss (Fitting, Working Electrode), Coated Titanium (Counter Electrode), PEEK • Sensitivity: First bacterial layer • Max. Pressure: Determined by Tri-Clamp Used (1.5" TC Connection) • Operating Temperature: -10 °C to 150 °C • Temperature Required for Biofilm Monitoring: 2 °C to 40 °C • Required Oxygen Levels: > 1 ppm • Required Conductivity Levels : >10 μS/cm • Power Supply: 12-24 V DC, 500 mA • Data Communication : 4-20 mA or RS485/MODBUS RTU (Cannot use both) • Wiring: 6-wire cable. 2 wires used for power supply, 2 for RS485/MODUS communication, 2 for 4-20 mA data transmission • Software - Minimum System Requirements (RS485/MODBUS): PC with Windows XP/7/8/10, 1 GHz CPU, 512 Mb Ram, 200 Mb of free space on hard drive, RS485 serial interface or USB port (for USB-RS485/MODBUS converter)