TD-0003_SaniCab® P_Technical Datasheet_Rev8.0

10.3 RECIRCULATED HEATING There are three (3) options for changing the design of the base system’s method of heating the recirculation system’s cleaning solution: • Heat Exchanger – Steam • Electric Heater to Maintain Temperature • No Recirculated Heating The Heat Exchanger – Steam option removes the direct steam injection assembly in the standard system and replaces it with a shell and tube heat exchanger system for heating of the recirculation system. • Steam Piping Design ο Material: Carbon Steel SCH 80 pipe ο Weld Finish: High-temperature black painted welds and pipe (carbon steel pipe) The following components are added: • One (1) Steam heat exchanger with the following specifications: ο Duty: See Table 8: Steam Heat Exchanger Data ο Material: 304ss wetted surfaces / 304ss non-wetted surfaces ο Design: Shell & tube ο Rating: ASME ο Connections: Tri-clamp (sanitary) / NPT (non-sanitary) • One (1) Steam supply piping manifold constructed of non-insulated SCH 80 carbon steel socket weld pipe and fittings ο One (1) Y-strainer with blow-off valve ο One (1) Pressure gauge with 0-100 psi range, anti-siphon, and manual block valve ο One (1) Modulating control ball valve with pneumatic actuator and positioner ο One (1) Pressure relief valve (set at 125 psi) piped to the floor ο One (1) Vacuum breaker with brass construction • One (1) Condensate return piping manifold constructed of non-insulated SCH 80 carbon steel socket weld pipe and fittings

ο One (1) Manual ball valve for condensate drain ο One (1) Float and thermostatic steam trap • Condensate drip-leg by-pass manifold ο One (1) 0.5” Thermodynamic steam trap

ο One (1) 0.5” Swing check valve (bronze) in the condensate drip-leg by-pass manifold to prevent back flow when heating ο One (1) 1.5” Swing check valve (bronze) in the condensate manifold to prevent back flow when the system is idle

TABLE 8: STEAM HEAT EXCHANGER DATA

Skid Dimensions

Approx. Heat-up Time 68 °F to 140 °F (min)

Connection Size / Type

Connection Size / Type (Condensate Return)

Chamber Model

Load (lbs/hr)

(Steam Supply)

W"

D"

H"

P335 P365 P565

20 20 20

50 50 50

70 70 70

3 6 5

2" FLG 2" FLG 2" FLG

1.5" FLG 1.5" FLG 1.5" FLG

966 966 966

NOTE: • Sani-Matic standard is to design steam heating applications for 50 psi steam. Installations where the plant supplied steam is more than 50 psi may require a pressure reducing valve (PRV) installed (by others).