10.12 CHAMBER STRAINING The Automated Conveyorized Mesh Screen option replaces the manual perforated screen tray strainer with a conveyorized strainer to the sump of the washer. This option is used for applications with large soil loads that do not dissolve in the initial rinse water. The soils are rinsed off the parts during the first rinse phase of the cleaning cycle and automatically conveyed out of the chamber. The following major components make up the strainer:
• Mesh strainer screen (1000 micron mesh size) • 304ss strainer cover / guard with EPDM flap • 1 / 3 HP painted steel gear drive
See Section 12.2.4 Conveyorized Strainer Catch Pan for an accessory to catch soils from the automated strainer. Customer supplied waste totes can also be used to catch accumulated soils. The maximum particulate height is 1 1 / 8 ” due to external opening gap.
Automated Conveyorized Mesh Screen - Outside View
10.13 RECIRCULATED STRAINING The Y-Strainer – Wedgewire Element recirculated straining option adds an additional level of straining that further protects the recirculation system’s rotary spray hubs and spray nozzles from excessive soil loading. The following components are added to the
recirculation line downstream of the pump outlet: • One (1) Y-Strainer body with tri-clamp connections • One (1) Strainer insert (0.010” wedgewire) • One (1) Pressure gauge with 0-100 psi range
ο The pressure gauge is added so that the system contains two (2) pressure measurement indicators; one (1) upstream and one (1) downstream of the Y-Strainer. This allows for monitoring of the soil loading of the Y-Strainer through the differential pressure.
10.14 DISTRIBUTION ZONES The Chamber Rotary Sprays & Active Coupler option adds an active coupler distribution zone to the cabinet washer supply. This zone, when combined with an active rack, allows for cleaning of hard-to-access areas of container-like parts (e.g., jugs with narrow openings), the insides of long narrow parts (e.g., hoses), and other complex parts. The following components are added to the system: • Two (2) Pneumatic butterfly valves for zoning of the chamber sprays and the active coupler • One (1) Active coupler, spring-loaded • Associated process piping • System automation for added zone valves