4. Utility Requirements
The following are the minimum required utilities for the proper operation of the system
Table 1: Utility Requirements
Chamber Model
P335
P365
P565
Connection Size / Type
15" TC
15" TC
15" TC
Water - Supply
Working Volume (gal)
18
40
26
Connection Size / Type
30" TC
40" TC
40" TC
Drain
Flow Rate (gpm)
40
70
70
Connection Size / Type
15" FLG
15" FLG
15" FLG
Steam
Load @ 50 psi (lbs / hr)
790
790
790
Connection Size / Type
05" FNPT
05" FNPT
05" FNPT
Instrument Air
Volume (scfm)
5
5
5
Disconnect Size @ 460V AC (amps)
Electrical
30
60
60
Chemical
Connection Size / Type
025" FNPT
025" FNPT
025" FNPT
Connection Size / Type
5" x 85" FLG
5" x 85" FLG
5" x 85" FLG
Exhaust
Volume (acfm)
900
900
900
Humidity (% RH)
100 Max
100 Max
100 Max
NOTE: • Drain and exhaust temperatures are variable based on the cycle • Drain connection is typically 3-6" off floor, depending on model and configuration • Water Supply − Two (2) Water supply connections are provided − Working volume is the typical chamber water volume required for operation of a single cleaning cycle phase Rack design and configuration may impact the working volume requirements − Total water volume required per cleaning cycle is dependent on cleaning cycle phases – eg, a typical P335 cleaning cycle may contain four (4) fills, requiring 72 gallons per cleaning cycle − Varying water supply temperatures are accommodated • Chemical − Three (3) chemical ports located in the recirculation pump suction manifold are provided − Non-foaming and non-chlorinated chemicals are required for effective and safe operation of the SaniCab − Backflow prevention is required (provided by others) − See Section 10.13 Chemical Addition for a complete system solution to add chemical pump and container assemblies to the system − See Section 12.2.2 Chemical Pump Assembly, Wilden P.025 for a loose ship chemical pump system solution
sanimatic.com 7
TD-0003.5
Powered by FlippingBook