• Electrical Components − One (1) 1 HP motor starter for the dryer supply fan − One (1) Controller for dryer electric heater
Adding service panels to the dryer system provides separation of the dryer maintenance space to the surrounding room interfaces The Dryer System – Front Service Panels (Single Door) includes service panels on the front load/unload side of the system The Dryer System – Front and Back Service Panels (Double Door) option includes service panels on both the front load side and back unload side of the system
Table 7: Dryer Utilities
Dryer - Air Supply
Dryer Air - Exhaust
Chamber Model
Connection Size / Type
Volume (CFM)
Temperature (°C)
Humidity (%RH)
Connection Size / Type 60" Flange 80" Flange
Volume (CFM)
Temperature (°C)
Humidity (% RH)
SP333 SP443 SP444
N/A N/A N/A
400 400 400
20 20 20
70 70 70
400 400 400
82 82 82
100 100 100
80" Flange
NOTE: • All dryer system options add 23" height to the overall system dimensions (see Table 2: System Size ) • All dryer system options add 900 lbs for model SP333 and 1,000 lbs for the SP443 and SP444 models to the system weight (see Table 2: System Size ) • If selecting a dryer system option with service panels, the appropriate option should be chosen based on the selected door configuration • Air supply for the dryer is drawn through the pre-filter from surrounding maintenance space air
10.5 Recirculated Heating The Steam option removes the electric heater from the base system and replaces it with a steam coil system for heating of the circulation system The following components are added: • One (1) Steam coil with the following specifications: − Duty: See Table 8: Steam Heating Data − Material: 316Lss − Insulation: None − Design: In-line coil − Rating: Non-ASME − Connections: 60" Tri-clamp (process), 05" tube - compression fittings (utility) • One (1) 60" Hygienic steam coil housing assembly • One (1) Steam supply piping manifold constructed of SCH80 carbon steel socket weld pipe and fittings − One (1) Y-Strainer with blow-off valve − One (1) Pressure gauge with manual ball valve for isolation − One (1) Modulating control v-ball valve with pneumatic actuator and positioner − One (1) Steam pressure relief valve (set at 125 psi) piped to the floor − One (1) Vacuum breaker with brass construction • One (1) Condensate return manifold constructed of SCH80 carbon steel socket weld pipe and fittings − One (1) Manual ball valve for condensate drain − One (1) Float and thermostatic steam trap − One (1) Check valve for condensate return isolation • One (1) Condensate bypass manifold constructed of SCH80 carbon steel socket weld pipe and fittings
− One (1) Balanced pressure thermostatic steam trap − One (1) Check valve for condensate return isolation