TD-0008_SaniCab® PC_Technical Datasheet_Rev3.0

10.6 Supply Pressure Monitoring The Pressure Transmitter option replaces the pressure gauge in the recirculation system downstream of the angle-line strainer with a diaphragm type pressure transmitter with local indication The pressure transmitter has a 0-100 psi range and 4-20 mA output This option adds system monitoring and alarming capabilities for the recirculation system pressure 10.7 Supply Flow Monitoring The Flow Switch option adds an insertion probe style flow switch to the recirculation system The flow switch has a discrete output This option adds system monitoring and alarming capabilities for the recirculation system flow, based on the discrete output The Flow Meter option adds a turbine flowmeter with local indication to the recirculation system The flow meter has a 25-250 GPM range and 4-20 mA output This option adds system monitoring and alarming capabilities for the recirculation system flow 10.8 Chemical Control The Wash & Rinse Conductivity Control option replaces the standard wash conductivity sensor, which can read between 200µS/cm and 1,000mS/cm with a more sensitive sensor with dual ranges that can read between 1µS/cm to 200mS/cm The two range options are as follows: • Wash Conductivity – used for controlling chemical wash steps − 02 ms/cm – 200 ms/cm Range • Rinse Conductivity – used for ensuring final rinse conditions (low conductivity) are met − 1 µS/cm – 50 mS/cm Range 10.9 Reporting These options can replace the standard No Reporting option for recording the critical cleaning cycle parameters of the SaniCab PC: • SaniTrend ® Local

• SaniTrend ® Cloud CR Essentials • SaniTrend ® Cloud CR Insights • SaniTrend ® Cloud CR Insights+

The SaniTrend ® Local reporting option provides a means of storing, retrieving, and printing the SaniCab PC cleaning cycle data with a desktop PC

The PC-based reporting system is installed on a desktop PC and interfaces with the Sani-Matic cleaning system’s Programmable Logic Controller (PLC) to collect and store cleaning wash cycle data An authorized user logs into the SaniTrend PC to select recipes by date or time through a recipe selection screen When the recipe is selected the user can view the recipe data as Recipe Chart Data, Recipe Charts, and Recipe Events screens The system provides easy access to the cleaning cycle data generated The data collected is recorded, stored, and the information Sample SaniTrend Local Printout

is printable Reference the SaniTrend Local brochure for more information SaniTrend ® Cloud Cleaning Records (CR) Essentials is an online data software that provides automated, secure data acquisition and reporting of critical cleaning cycle information, along with system operational efficiency insights Users, having a facility user login account, access SaniTrend Cloud via a web browser The Essentials edition of SaniTrend Cloud includes the following: • (1) Industrial PC / Edge Device to buffer and transmit data to the cloud portal • Connectivity software installed on the industrial PC / edge device to allow it to connect to the PLC • SaniTrend Cloud online portal access • Product features: − Cloud Access to Data (Phone, Tablet, Mobile) − Unlimited Users

OEE Reports

Cleaning Cycle Report

− Store and Forward Capabilities − Cloud Documentation Storage − Live Dashboard Views

sanimatic.com 19

TD-0008.3