Settings
Step 3: Press the “Advanced Settings" button, then press the “Pump Rotation Test” button. At any time, you can press the “back” button to return to the previous screen.
Back
Advanced Settings
Pump Rotation Test
Step 4: Press the “Start Test” button. This will activate the pump motor for 0.5 seconds, enough to allow the fan to spin. Quickly observe the back of the motor once the “Start Test” button is pressed. The fan should be spinning CLOCKWISE. If the fan is NOT spinning clockwise, an ensure proper LOTO procedures have been followed and consult an electrician to switch the phases.
CleanWorx ™ CyClean ™ 35 IOM
Powered by FlippingBook