Step 4: After the CyClean has been leveled at the correct height, with the Sanitation Cart inserted, confirm the CyClean door can be closed without any interferences. If the Cyclonejet 360° Spray Arm was removed, it can be reinstalled at this time. Likewise, if any zip-ties or other fastening methods were used to keep the spray arm from moving, they should be removed at this time. Depending on what is being loaded onto your Sanitation Cart, additional adjustments might need to be made. See section 6.4) Adjusting a Sanitation Cart for more details.
4.3) Initial Start-up Pre-Check
Before initial start-up, check each utility supply to verify complete and secure connections are made and that the connections meet the requirements listed in section 2.2) Utility Connections . Common issues include, but are not limited to: • The water fill valve is either not connected, the water supply line is restricted, or the water supply line is not sized appropriately. • The drain port valve is not connected, the drain line is restricted, or the drain line is not sized appropriately. • The exhaust port is not connected, the exhaust line is restricted, or the exhaust line is not sized appropriately. • If used, chemical line(s) are not connected to the chemical port(s), the proper electrical connections are not made, or the appropriate accessories have not been enabled in the HMI screen. • Fuses, overload protection, and circuit breakers are not installed or properly sized. • Electrical service is either not connected to the control cabinet or is not the proper voltage defined in section 2.2) Utility Connections .
Using the proper tools, ensure the incoming power to the panel is the correct voltage at the disconnect before turning on the main disconnect.
21
Powered by FlippingBook