Typically, this Step is used as a “final rinse”, intended to be a relatively short runtime to remove any residual chemicals.

Operations: (1) Add water.

(2) Heat water to temperature Setpoint. (3) Run the pump for runtime Setpoint. (4) Depending on operator selection and recipe steps: Drain used water. If this is the final step, the water will either be held in the sump to be used as the pre-rinse water in the next cycle or will be drained based on the operator's selection. If this is NOT the final step, the water is drained after this step is complete. There are three (3) main uses for the sanitizer wash Step: 1. As intended, apply a leave-on chemical sanitizer. The sanitizer will only be applied for the specified amount of time. (Typical) 2. Skipping this step as only a single final rinse is needed. (Typical) 3. As a secondary final rinse. (Uncommon) Operations: (1) Water is added. (2) Heat water to temperature Setpoint. (3) Add sanitizer for time Setpoint. (4) Run the pump for runtime Setpoint. (5) Drain used water. The water is always drained after this step as a safety precaution.

Rinse

Sanitizer

Regardless of the configuration of the Wash Recipe, once the desired Steps have been completed, the CyClean exhaust will activate for a minimum of 30 seconds. The purpose of this is to reduce the humidity inside the CyClean before an operator opens the door. After the exhaust step, the CyClean Wash Recipe will be complete.