E05 – Empty Hold Watchdog Possible Cause
Solution
a) Drain valve has been open for longer than the respective setpoint value and the level sensor is still detecting water
- Adjust setpoint through HMI - Repair or replace level sensor - Repair or replace drain valve - Inspect drain line for blockage
E06 – High Temperature Alarm Possible Cause
Solution
a) Water temperature has exceeded respective setpoint value
- Adjust setpoint through HMI - Repair or replace temp sensor - Lower incoming water supply temperature
E07 – Temperature Hold Watchdog Possible Cause a) Heater has been active longer than respective setpoint value
Solution
- Repair or replace temp sensor - Adjust setpoints through HMI - Inspect heating elements then clean or replace as necessary
E08 – Heater High Temp Safety Failure Possible Cause a) Heater internal temperature has exceeded the physical safety limit
Solution
- Repair or replace heater - Lower incoming water supply temperature
E09 – Heater Contact Alarm Possible Cause a) Internal heater wiring fault E10 – Recirculation Pump Failure Possible Cause a) Pump motor is not activating
Solution
Repair or replace electric heater
Solution
- Ensure motor is wired correctly - Check fuses, breakers, & overloads