DM-0018_CleanWorx™ CyClean™ 35_IOM_Rev1.3

10) APPENDIX

10.1) Factory Set System Setpoints and Limits

The table below contains the default system Setpoints with their respective minimum and maximum allowed value as addressed in section 5.2) Advanced Settings . These Setpoints are used across all Wash Recipes. System Setpoint Units Default Value Min Value Max Value Notes 1 High Temperature Setpoint °F 195 50 210 2 Sump Fill Time Setpoint sec 18 0 60 3 Sump Drain Time Setpoint sec 78 60 180 4 Add Water Pulse Time Duration sec 5 2 60 5 Chemical Burst On Time Setpoint sec 5 2 9999 6 Chemical Burst Off Time Setpoint sec 5 2 9999 7 Control Setpoint Addition n/a 5 2 9999 8 Temperature Watchdog Timeout sec 3600 0 9999 9 Conductivity Watchdog Timeout sec 300 0 9999 10 Sump Fill Watchdog Timeout sec 300 0 9999 11 Sump Drain Watchdog Timeout sec 300 0 9999 12 Temperature Minimum Scale °F 30 0 9999 13 Temperature Maximum Scale °F 230 0 9999 14 Conductivity Minimum Scale mS/cm 0 0 9999 15 Conductivity Maximum Scale sec 100 0 9999