TD-0003_SaniCab® P_Technical Datasheet_Rev8.0

SANICAB ® P CABINET WASHER

Usable chamber dimensions vary by model size, as shown below in Table 4: Usable Space Dimensions . The usable space dimensions identify the practical space available for parts loading on the rack.

TABLE 4: USABLE SPACE DIMENSIONS

Usable Space Dimensions

Chamber Model

Usable Space Volume (ft 3 )

Usable Space Dimensions

W"

D" 35 73 71

H" 65 65 54

P335 P365 P565

30 30 56

39 82

124

TABLE 5: PUMP DATA 7.2 RECIRCULATION SYSTEM The recirculation system draws cleaning solution from the chamber sump and distributes it to the chamber’s rotary sprays. It is self-contained within the system’s footprint. • Recirculation system components ο One (1) pneumatic butterfly valve for system draining ο One (1) 3.0" tri-clamp chemical injection port » Tri-clamp cap contains three (3) 0.25" FNPT ports for chemical injection ο One (1) Centrifugal supply pump » 316Lss wetted materials with EPDM elastomers » 3500 rpm, TEFC motor ο One (1) Temperature transmitter with 30-230 °F range and 4-20 mA output ο One (1) Pressure gauge with 0-100 psi range ο One (1) Sanitary piping manifold Chamber Model HP Pump Rating gpm P335 P365 10 20 200 400 P565 15 300

psi

43 50 45

NOTE: • Pump rating data is based on Ampco pumps. Ratings may change for other pump manufacturers.

Chamber Model 7.3 HEATING SYSTEM Steam is injected directly into the chamber sump to heat the recirculating cleaning solution. The steam assembly contains the following components: • One (1) Steam sparger assembly ο Duty: See right Table 6: Direct Steam Injection Performance ο Material: 316ss with welds color cleaned ο Insulation: None ο Design: Direct steam injection ο Rating: Non-ASME ο  Connections: Tri-clamp (sump connection), MNPT (steam connection) • Steam supply piping manifold constructed of uninsulated SCH 40 304ss socket weld pipe and fittings ο One (1) Pressure gauge with 0-100 psi range anti-siphon and manual block valve P335 P365 P565

TABLE 6: DIRECT STEAM INJECTION PERFORMANCE

Approx. Heat-up Time 68 °F to 140 °F (min)

2 3 3

NOTE: • For direct steam injection, the SaniCab P requires a steam pressure of 50 psi or less. Installations where the plant supplied steam is more than 50 psi, a pressure reducing valve (PRV) is required (by others).

ο One (1) Steam strainer with manual blow-down valve ο One (1) On/off ball valve with pneumatic actuator ο One (1) Check valve

10

SANICAB® P CABINET WASHER TECHNICAL DATASHEET

Powered by