TD-0002_PharmaCab® SP_Technical Datasheet_Rev4.0

11.2 System Additions 11.2.1 Passivation – System

Sani-Matic will provide factory passivation of all solution contact surfaces on the system The citric acid passivation is performed and documented per Sani-Matic’s standard procedure (SOP-MA005), which adheres to ASTM A 967-05 requirements A certificate of passivation (SMI-LOG-007) is provided upon completion 11.2.2 Hydrostatic Test Under this System Addition option, Sani-Matic will provide a hydrostatic test of the system’s piping The hydrostatic test is performed per Sani-Matic’s standard procedure (SOP-QA007) based on ASME guidelines A certificate of hydrostatic testing (SMI-LOG-002) is provided upon completion 11.2.3 Riboflavin Test – Chamber This is a customer-witnessed riboflavin spray coverage

test for areas of the chamber which have the potential to drip onto a rack Sani-Matic follows internal SOPs that are developed using ASME BPE Nonmandatory Appendix L as a guideline A test report (SMI-LOG-004) is provided, documenting the system test parameters 11.2.4 Riboflavin Test – Rinse Tank This is a customer-witnessed riboflavin spray coverage test for the final rinse tank Sani-Matic follows internal SOPs that are developed using ASME BPE Nonmandatory Appendix L as a guideline A test report (SMI-LOG-004) is provided, documenting the system test parameters

NOTE: • This accessory can only be selected if the Single Pass Final Rinse system option is selected

11.2.5 Borescope Inspection Video - System The system's process piping hygienic welds are borescope inspected and a DVD video record is provided 11.2.6 Seismic Zone Anchorage Calculations Seismic calculations are performed for the anchorage of the system to the plant floor The services are performed by an engineer certified in the state where the system will be installed in and includes documentation on the calculations and a licensed engineer's stamp

NOTE: • Due to the unique seismic zone considerations for each installation, modifications to the system design and footprint may be required, which may incur additional charges

11.2.7 Barrier Flanges Eight pieces of L-shaped 16 ga, 32 μin Ra, 304ss barrier flanges (profile dimensions: 15" x 25" x 11' L) are shipped loose for both the load side and unload side to allow the exterior of the washer to be sealed to the walls, floor, and ceiling to reduce airflow between areas of different pressurization The barrier flange design requires adhesive to attach the flanges to the cabinet washer frame and paneling

sanimatic.com 23