CLEAN-IN-PLACE SYSTEMS
OPERATING REQUIREMENTS
TACTaFlow S60 / D60 Description
Requirement
Connection
Water #1 – Purified Water
20 gpm @ ≤ 40 psi, 20 °C - 80 °C
1.5” TC
Water #2 – WFI
20 gpm @ ≤ 40 psi, 20 °C - 80 °C
1.5” TC
Drain
≤ 60 gpm @ 20 °C - 80 °C
2.0” TC
Instrument Air
10 cfm @ 90 psig
0.5” FNPT
Steam Supply
450 lbs/hr @ 50 psig
1.0” Flange
Condensate Return
450 lbs/hr
1.0” Flange
30 Amp @ 460V AC, 3PH, 60 Hz (100A with Electric Heater option)
Electrical Supply
--
CIP Return (CIPR) (Optional)
10-60 gpm @ 30 psig
2.0” TC
TACTaFlow S45
Description
Requirement
Connection
Water #1 – Purified Water
20 gpm @ ≤ 20 psi, 20 °C - 80 °C
1.0” TC
Water #2 – WFI
20 gpm @ ≤ 20 psi, 20 °C - 80 °C
1.0” TC
Drain
≤ 45 gpm @ 20 °C - 80 °C
2.0” TC
Instrument Air
10 cfm @ 90 psig
0.5” FNPT
100 Amp @ 460V AC, 3PH, 60 Hz (30A with Steam Heat option)
Electrical Supply
--
NOTE: Operating requirements may vary based on selected options.