4. Utility Requirements
The following are the minimum required utilities for the proper operation of the system
Table 2: Utility Requirements Chamber Model
PC335
PC365
Connection Size / Type
15" TC
15" TC
Water - Supply
Working Volume (gal)
35
60
Connection Size / Type
30" TC
30" TC
Flow Rate (gpm)
Drain
55
55
Time to Completely Drain (s)
40
70
Connection Size / Type
05" FNPT
05" FNPT
Instrument Air
Volume (scfm)
5
5
Disconnect Size @ 460V AC (amps)
Electrical
100
100
Chemical
Connection Size / Type
075" TC
075" TC
Exhaust
Connection Size / Type
12" FLG
12" FLG
NOTE: • Drain and exhaust temperatures are variables based on the cycle
• Exhaust connection is shared with the dryer system (see Section 10.3 Dryer System ) Due to temperature variations within the chamber, this connection must not be blocked • Water Supply: − Two (2) Water supply connections are provided − The working volume is the typical chamber water volume required for the operation of a single cleaning cycle phase Rack design and configuration may impact the working volume requirements − Total water volume required per cleaning cycle is dependent on cleaning cycle phases – eg, a typical PC335 cleaning cycle with re-used final rinse for the next pre-rinse may contain three (3) fills, requiring 105 gallons per cleaning cycle − Varying water supply temperatures are accommodated • Chemical: − Three (3) chemical ports located in the recirculation pump suction manifold are provided − Non-foaming and non-chlorinated chemicals are required for effective and safe operation of the SaniCab PC − See Section 12.2.2 Chemical Pump Assembly – Diaphragm Pump for a loose ship chemical pump system solution − See Section 12.2.3 Chemical Pump Assembly – Electronic Metering Pump for a loose ship chemical pump system solution with finer chemical controls − See Section 12.2.4 Chemical Delivery Skid – Diaphragm Pump for a complete chemical system solution skid to add to the system − See Section 12.2.5 Chemical Delivery Skid – Electronic Metering Pump for a complete chemical system solution skid with finer chemical controls
sanimatic.com 9
TD-0008.3
Powered by FlippingBook