4. Utility Requirements
The following are the minimum required utilities for the proper operation of the system.
Table 1: Utility Requirements Description
Requirement
Connection
Water - Supply
110 gpm @ 30 psi
1.5" TC
Water - Drain (Suction Manifold)
Up to 50 gpm
2.5" TC
Water - Drain (Return Manifold)
Up to 120 gpm
2.0" TC
Instrument Air
10 scfm @ 90 psi
0.5" NPT
Electrical
30 Amp @ 460 V AC, 3 PH, 60 Hz.
N/A
CIP Supply
N/A
2.0" TC
CIP Return
N/A
2.0" TC
NOTE: • Drain temperatures are variable based on the cycle.
5. Physical Size & Layout
The following is the approximate overall size of the equipment.
Table 2: System Size
System Dimensions
Description
W"
D"
H"
Standard System Model
74"
39"
75"
NOTE: • System dimensions and system weight listed are for standard options – certain options or combination of options can increase the overall size of the equipment.