a. Date and Time
Note (1)
Note (2)
Note (5)
Note (3)
V
Note (4)
Note (6)
Notes The date and time can be set manually or by specifying a remote NTP server. By default, the time is set by the time.windows.com NTP server from Microsoft. This can be changed to any internet NTP server (if IT has allowed communication), or a local NTP server on the network. If a firewall is in place, UDP port 123 shall be open to allow the clock synchronization through the SNTP protocol. 1. The time synchronization mode can be set to automatic or manual by selecting the appropriate radio button. 2. When time synchronization is set to automatic (“Auto”), the remote NTP server can be specified by IP address or DNS name. 3. The time zone can also be specified and should be set up to be the local time zone where the DPR is installed. 4. The date and time can be entered manually if the time synchronization mode is set to “Manual”. 5. If using a Local NTP server, the “LAN” checkbox must be ticked. Unchecking the box will disable local NTP interfaces as the DPR HMI panel has 1 Ethernet port (LAN). 6. Any changes to the date and time will require restarting the DPR’s HMI panel. When necessary, a reboot message appears and the “Reboot” button is enabled. Press the reboot button to commit the changes and restart the HMI panel.
DM-0013_Digital Process Recorder (DPR)_IOM_Rev4.0
19 of 76
Powered by FlippingBook