TD-0001_COP Parts Washers_Technical Datasheet_Rev6.0

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER Technical Datasheet

MODEL NUMBER KEY

EXAMPLE MODEL #: RW06-431NP2S-ASV-413010 3 5 7 9 1 2 4 6 8 1011 12 1314 151617181920

TANK SERIES UW 20" W x 18" D RW 24" W x 22" D PW 24" W x 30" D TW 30" W x 24" D TANK LENGTH 04 4' L 06 6' L 08 8' L

ELECTRICAL UTILITY 4

1, 2

15

460V AC, 3PH, 60 Hz (Standard)

CONTROLS 0

16

No Controls (Standard)

1 2

Disconnect with Start / Stop Buttons

Disconnect with Start / Stop Buttons and Electric Heater Controller

3, 4

CONDUIT 0

17

No Conduit (Standard)

1 2 3 4

PVC Conduit

10 10' L 12 12' L

PVC Conduit - For Electric Heater

Stainless Steel Conduit

Stainless Steel Conduit - For Electric Heater

WETTED MATERIALS OF CONSTRUCTION 4 304 Stainless Steel (Standard)

5

CHEMICAL CONTROL 0

18

No Chemical Control (Standard)

TANK INTERIOR SURFACE FINISH 3 32 Ra with Welds Ground (Standard)

6

1

Conductivity Analyzer

TANK COVER 0

REPORTING 0

7

19

No Cover (Standard)

No Reporting (Standard)

1

Hinged Cover

1 2

1-Pen Chart Recorder, Temperature Recording

2-Pen Chart Recorder, Temperature & Conductivity Recording

DRAIN BOARD / PIPE RACKS N No Drain Board or Pipe Racks (Standard) D Drain Board with Pipe Racks

8

C Digital Process Recorder (DPR) Charts

FILL PORTS 0

20

No Fill Ports (Standard)

STATIONARY / PORTABLE S

9

1

Water and Chemical Fill Port

Stationary with Adjustable Feet (Standard)

P

Portable with Casters

JET MANIFOLDS 1

10

Side Jet Manifolds (Standard) Side Jet & End Jet Manifolds

2 3 4

Side Jet & Hose Cleaning Manifolds

Side Jet, End Jet, & Hose Cleaning Manifolds

HEATING S

11

In-Tank Steam Sparger (Standard) Electric Heater to Maintain Temperature

E

N No Heating Method 12, 13 PUMP & MOTOR TYPE

AT Ampco Pump - TEFC Motor (Standard) AW Ampco Pump - Washdown Motor AS Ampco Pump - Stainless Steel Washdown Motor

VALVE TYPE V

14

VNE Butterfly (Standard)

2

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER TECHNICAL DATASHEET

TABLE OF CONTENTS

1.Introduction . . . . . . . . . . . . . . . . . . . . . . ............. 4 2.Applications . . . . . . . . . . . . . . . . . . . . . . ............. 4 3.Construction.................................. 4 3.1 Certifications and Classifications .. .. .. .. .. ......... 4 3.2StructuralFraming............................ 4 3.3 Sanitary Piping . . . . . . . . . . . . . . . . . . . . ........... 5 3.4SteamPiping . . . . . . . . . . . . . . . . . . . . ............ 5 4.UtilityRequirements............................. 6 5.PhysicalSize . . . . . . . . . . . . . . . . . . . . . . ............. 7 6. Component Manufacturers . . . . . . . . . . . . . . . . ......... 8 7. General Equipment Description . . . . . . . . . . . . . . ........ 8 7.1Tank.. .. .. .. .. .. .. .. .. .. .. .. .............. 8 7.2 Circulation System . . . . . . . . . . . . . . . . . . .......... 9 7.3 Heating System . . . . . . . . . . . . . . . . . . . ........... 9 8.Testing.. .. .. .. .. .. .. .. .. .. .. .. ............... 10 9. Documentation . . . . . . . . . . . . . . . . . . . . . ............ 10 10. Product Options . . . . . . . . . . . . . . . . . . . . ........... 11 10.1TankCover ...............................11 10.2 Drain Board / Pipe Racks . . . . . . . . . . . . . . ......... 11 10.3 Stationary / Portable . . . . . . . . . . . . . . . . ......... 11 10.4 Jet Manifolds . . . . . . . . . . . . . . . . . . . ........... 12 10.5Heating.. .. .. .. .. .. .. .. .. .. .............. 12 10.6 Pump & Motor Type . . . . . . . . . . . . . . . . .......... 13 10.7Controls.................................13 10.8Conduit. . . . . . . . . . . . . . . . . . . . . ............. 14 10.9 Chemical Monitoring . . . . . . . . . . . . . . . . ......... 14 10.10 Reporting . . . . . . . . . . . . . . . . . . . . ............ 14 10.11FillPorts . . . . . . . . . . . . . . . . . . . . . ............ 15 11. Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 11.1 Chemical Pump Assembly, Wilden P1 . . . . . . . . . . . . . . 16 11.2EndMountedDrainTable . . . . . . . . . . . . . . . . . . . . . . 17 11.3 Turn Over Package (TOP) - Hard Copy . . . . . . . . ...... 17 11.4 WiFi Connection Package . . . . . . . . . . . . . . ........ 17 11.5COPPartsBaskets..........................18

3

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER

1 INTRODUCTION

Technical information provided within this document is for Sani-Matic’s Clean-Out-of-Place (COP) Parts Washer Configured Design Offering (CDO) serving the food and beverage industry. Alternative Design Requests (ADR) for variations in construction, options, orientation, or other technical requirements should be directed to the Sani-Matic sales team for custom quoting and engineering. The system details described in Sections 2 through 9 apply to all models and their standard options, which are designated with “(Standard)” in the model key. The selectable product options within the model key are described in Section 10 . Accessories such as chemical pumps and COP Parts Baskets are described in Section 11 . Sani-Matic’s Clean-Out-of-Place (COP) Parts Washers clean food & beverage industry process parts that cannot be cleaned in place and require disassembly for thorough cleaning. Typical process parts cleaned include pipes, hoses, fittings, clamps, gaskets, and valves. 2 APPLICATIONS

Fittings, Gaskets & Clamps

Miscellaneous Parts

Hoses & Pipes

Valves

3 CONSTRUCTION

3.1 CERTIFICATIONS AND CLASSIFICATIONS • Electrical Area Classification: Non-hazardous • Electrical Certification: UL 508A • Seismic Design: None • Industry Standards: Authorized to provide Canadian Registration Numbers (CRN), helps meet FSMA and HACCP Sanitation Preventive Controls • System Certifications: None 3.2 STRUCTURAL FRAMING • Material: 304ss round tubing • Surface Finish: Bead blasted • Fully welded, single-piece construction • Feet for locating the equipment at a fixed location in the facility. The feet are adjustable allowing the equipment to be leveled on uneven or sloped surfaces. Models with lengths 8’ and less have four (4) feet, while models 10’ and longer have six (6) feet.

4

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER TECHNICAL DATASHEET

3.3 SANITARY PIPING • Material: 304ss sanitary tube & fittings • Surface Finish: 32 µin Ra ID • Weld Finish: As welded ID / weld color removed OD (bead blast as part of tank construction) • Welding: Welds are performed manually per AWS D18.1/D18.1M standards (latest edition). The weld interior is argon gas purged. • Slope: None • Connections: Tri-clamp fittings with single hinged heavy-duty clamps with wing nuts • Gasket Material: EPDM • Valves: Butterfly type with 304ss body, EPDM seat, and manual actuator

3.4 STEAM PIPING

• Material: Stainless steel pipe • Connections: NPT

5

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER

The minimum required utilities for the proper operation of the equipment are listed in Table 1 (below). If these utilities are not available, an Alternative Design Request may be required. 4 UTILITY REQUIREMENTS

TABLE 1: UTILITY REQUIREMENTS

Potable Water

Drain

Steam Supply

Electrical

Tank Series and Length

Approximate Time (min) for Full Tank Heat Up (55 °F to 140 °F)

Tank Fill (gallons)

Connection Size / Type

Load @ 50 psi (lbs/hr)

Connection Size / Type

Amps @ 230V AC

Amps @ 460V AC

UW04 UW06 UW08 UW10 UW12 RW04 RW06 RW08 RW10 RW12 PW04 PW06 PW08 PW10 PW12 TW04 TW06 TW08 TW10 TW12

62 91

2.0" TC 2.0" TC 2.5" TC 2.5" TC 2.5" TC 2.5" TC 2.5" TC 2.5" TC 3.0" TC 3.0" TC 2.5" TC 2.5" TC 3.0" TC 3.0" TC 3.0" TC 2.5" TC 2.5" TC 3.0" TC 3.0" TC 3.0" TC

377 377 742 742 742 377 742 742

12 12

0.50" FNPT 0.50" FNPT 0.75" FNPT 0.75" FNPT 0.75" FNPT 0.50" FNPT 0.75" FNPT 0.75" FNPT 1.00" FNPT 1.00" FNPT 0.75" FNPT 0.75" FNPT 1.00" FNPT 1.00" FNPT 1.00" FNPT 0.75" FNPT 0.75" FNPT 1.00" FNPT 1.00" FNPT 1.00" FNPT

7.0 7.0

3.5 3.5 6.0 6.0 9.0 6.0 6.0 9.0

120 149 178

8

12.0 12.0 18.0 12.0 12.0 18.0 36.0 36.0 12.0 18.0 36.0 36.0 36.0 12.0 18.0 36.0 36.0 36.0

10 10 12

95

140 185 230 275 138 203 267 332 397 132 195 257 319 382

9

12

1,272 1,272

9

18.0 18.0

11

742 742

9

6.0 9.0

13 10 13 15

1,272 1,272 1,272

18.0 18.0 18.0

742 742

9

6.0 9.0

13 10 12 15

1,272 1,272 1,272

18.0 18.0 18.0

NOTE: • All pumps are 230/460V AC dual rated. See Section 10.7 (Table 6) for disconnect size when selecting a controls package. • Drain temperature is variable based on COP cycle. • Steam Supply requirement is based on selection of the standard in-tank steam sparger. • Chemical requirement is dependent on cleaning process.

6

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER TECHNICAL DATASHEET

The approximate overall size of the equipment is listed in Table 2 (below). Both overall system dimensions and tank dimensions are listed, as well as the nominal tank volume. 5 PHYSICAL SIZE

TABLE 2: SYSTEM & TANK SIZE

System Dimensions (inches)

Tank Dimensions (inches)

Nominal Tank Volume (gallons)

Tank Series and Length

Width

Height

Length

Width

Height

Length

UW04 UW06 UW08 UW10 UW12 RW04 RW06 RW08 RW10 RW12 PW04 PW06 PW08 PW10 PW12 TW04 TW06 TW08 TW10 TW12

30 30 30 30 30 34 34 34 34 34 34 34 34 34 34 40 40 40 40 40

39 39 39 39 39 43 43 43 43 43 51 51 51 51 51 45 45 45 45 45

64 88

20 20 20 20 20 24 24 24 24 24 24 24 24 24 24 30 30 30 30 30

18 18 18 18 18 22 22 22 22 22 30 30 30 30 30 24 24 24 24 24

51 75 99

75

100 150 175 200 100 150 200 250 300 150 225 300 375 450 150 225 300 375 450

112 136 160

123 147

64 88

51 75 99

112 136 160

123 147

64 88

51 75 99

112 136 160

123 147

64 88

51 75 99

112 136 160

123 147

NOTE: • System dimensions listed are for standard options. The below images detail what the referenced dimensions are for an example configured design COP Parts Washer. Some product options add additional length, width, or height. See individual Product Options ( Section 10 ) for more details. • Design documents are sent for customer reference only. System piping and component configuration and tank layout are predetermined. If a custom layout or configuration is required, contact the Sani-Matic sales team for an Alternative Design Request COP Parts Washer quote.

SYSTEM LENGTH (ENCLOSURE DOOR OPEN)

PLAN VIEW

80° REF (COVER OPENING)

TANK LENGTH OVERALL LENGTH

OVERALL WIDTH

TANK WIDTH

OVERALL HEIGHT (W/ COVER OPEN)

TANK DEPTH

WORKING HEIGHT

STEAM SUPPLY

DRAIN

LEFT END VIEW

SIDE VIEW

RIGHT END VIEW

7

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER

The following are the manufacturers and models of the components used in the assembly of this equipment. Note that inclusion of certain components is dependent on the type of model options selected. 6 COMPONENT MANUFACTURERS

TABLE 3: COMPONENT MANUFACTURERS

Component

Manufacturer

Model

Pump

Ampco

AC

Valve, Butterfly

VNE

51 Series

Valve, Control (steam) Conductivity Sensor Conductivity Analyzer

Watson McDaniel / Trerice

W91 / D Series

Rosemount Rosemount Anderson ifm Efector ifm Efector Glo-Quartz

225

1056

Chart Recorder

AJ-300

RTD

TM TD

Temperature Transmitter

Electric Heater Chemical Pump

-

Wilden

P1

7 GENERAL EQUIPMENT DESCRIPTION

7.1 TANK • The tank is used to hold chemical wash solutions. All wash solutions are supplied to the tank’s jets or manifolds from the circulation system which draws the wash solutions from the tank. A sump strainer (bead blast finish) with 1 / 8 " perforations is located at the outlet of the tank to protect the pump and jets. The components being washed are placed in the tank, either loose or in baskets, and submerged in the wash solutions. • Design Features: ο Sides: 12 ga. ο Bottom Head: 12 ga. ο Insulation: None ο Overflow: None ο Design: Atmospheric ο Material: 304ss wetted surfaces ο Interior Finish: 32 µin Ra with welds ground smooth (strainer insert, tank trough, etc. may have other finishes) ο  Exterior Finish: Bead blasted • Standard equipped side jet manifolds create rolling agitation within the tank and are best for washing small parts and fittings. The jets are of a sanitary design being fully welded to the tank and the supply manifold. • The UW, RW, and TW series tanks have dual side jets, while the deeper PW series tanks have quad side jets.

8

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER TECHNICAL DATASHEET

7.2 CIRCULATION SYSTEM Utilizing a recirculated design, the circulation system draws wash solutions from the tank and supplies solution back to the tank through tank jets or manifolds. • One (1) Manual butterfly valve for drain. See Drain - Connection Size / Type within Table 1 for valve size. • One (1) Centrifugal pump: ο See Table 4 (below) for ratings ο 316Lss wetted materials with EPDM elastomers ο 3500 rpm motor ο TEFC Motor (standard)

TABLE 4: PUMP DATA

Pump Rating

Tank Series and Length

HP

gpm

psi

UW04 UW06 UW08 UW10 UW12 RW04 RW06 RW08 RW10 RW12 PW04 PW06 PW08 PW10 PW12 TW04 TW06 TW08 TW10 TW12

3.0 3.0 5.0 5.0 7.5 5.0 5.0 7.5

53 77

27 25 25 21 30 27 23 28 45 36 23 24 37 38 33 23 25 41 39 34

102 127 151

81

119 157 196 234 117 173 227 282 337 112 166 218 271 325

15.0 15.0

5.0 7.5

15.0 15.0 15.0

5.0 7.5

15.0 15.0 15.0

7.3 HEATING SYSTEM Steam is directly injected into the wash solution to elevate its temperature. See Table 1 for load requirements, approximate heat up time, and connection size and type. • Steam supply manifold ο One (1) Self-regulating control valve that is single seated and 316Lss wetted materials with bronze body components. The valve can control the wash solution temperature to a setting between 105–195 °F. Its thermobulb is mounted in a thermowell located in the tank’s sidewall. The thermowell has a bead blast, wetted finish. ο One (1) Steam mixer located in the tank. The mixer has a bead blast ID / color cleaned OD finish.

9

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER

8 TESTING The equipment will be factory tested to confirm its proper operation. Included in this are the below testing methods:

• Confirm that the equipment is complete and conforms to the drawings • Confirm that the safety systems operate correctly (if applicable) • Verify controls are wired properly (if applicable) • Functional testing of the control system (if applicable) • Wet testing of the pump

9 DOCUMENTATION

One (1) electronic copy of the documentation package is provided as standard. The documentation is provided in the English language and includes the following information: • Warranty information • Recommended spare parts (RSP) list • Device settings (if different than the vendor’s standard) • Piping & Instrumentation Diagram (P&ID) and Mechanical Bill of Material (BOM) • General Assembly (GA) drawing

• Electrical panel layout drawing (if applicable) • Electrical schematics drawing (if applicable) • Component manuals (as provided by the vendor)

10

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER TECHNICAL DATASHEET

10.1 TANK COVER The Hinged Cover option provides operator protection from the wash solutions in the tank, helping to contain vapors in the tank. The cover is hinged in the back and opens approximately 80°. For 4’- to 8’-length tanks, lift assist cylinders help reduce the effort to open the cover as well as hold it in the open position. On 10’- and 12’-length tanks, a counterweighted cover provides the same benefits. Both cover types are made of 304ss with weld color removed. NOTE: • This option cannot be used with the Drain Board with Pipe Racks option (see Section 10.2 ) or the End Mounted Drain Table accessory (see Section 11.2 ) when the cover is closed. • This option adds height (open cover) and width (cover hinge) to the system dimensions. 10 PRODUCT OPTIONS

Hinged Cover with Counterweight

Hinged Cover with Lift Assist Cylinders

10.2 DRAIN BOARD / PIPE RACKS Adding the Drain Board with Pipe Racks option allows for component draining and drying. The pipe racks are mounted to a drain board located along the rear edge of the tank. The drain board has upturned lips to contain residual wash solutions and is sloped to drain back to the tank. The pipe racks include four levels of rod supports to place long items (such as piping, hoses, or shafts) for draining or drying. Tanks 6’ and less have two (2) pipe racks, 8’ tanks have three (3) pipe racks, and 10’ and longer tanks have four (4) pipe racks. The drain board is integral to the tank while the pipe racks are capable of being removed. They are made of 304ss with weld color removed. NOTE: • This option cannot be used with the Hinged Cover product option (see Section 10.1 ). • This option adds approximately 28” in height and 3” in width to the system dimensions.

10.3 STATIONARY / PORTABLE Selection of the Portable with Casters option replaces the adjustable feet with casters allowing the equipment to be moved through the facility. The casters have 5” diameter non-marking polyurethane wheels, stainless steel swivel hardware, and foot brakes.

11

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER

10.4 JET MANIFOLDS The selection of adding End Jet Manifolds and/or a Hose Cleaning Manifold to the standard side jets adds additional cleaning actions as described below. Addition of the options adds the required manual butterfly valves to direct wash solutions to either the side jet manifolds, end jet manifolds, or hose cleaning manifold. Systems are sized to operate only one zone at a time and should not be run simultaneously. The standard side jet manifold is included when selecting either or both of these options.

• End Jet Manifolds create agitation lengthwise within the tank and are best for washing piping, hoses, and other long hollow items.

• The Hose Cleaning Manifold provides four (4) 1.5” tri-clamp connections (UW model tanks) or four (4) 2.0” tri-clamp connections (RW, PW, TW model tanks) internal to the tank, submerged in the fluid. Hoses and/or tubing can be attached to these connections providing flow of wash solutions through their interior.

NOTE: • The End Jet Manifolds option adds approximately 3” in length to the system dimension. 10.5 HEATING The direct steam injection heating system can be removed to provide No Heating Method or can be replaced with an Electric Heater to Maintain Temperature . • The Electric Heater to Maintain Temperature is sized to maintain temperature only (up to 170 °F) and is dependent on the system being filled with water already at this temperature. This does not include heat loss due to parts loading or room environment conditions.

TABLE 5: ELECTRIC HEATER DATA

Electric Heater Sizing kW

Tank Series and Length

Amp Draw (460V AC)

UW04 UW06 UW08 UW10 UW12 RW04 RW06 RW08 RW10 RW12 PW04 PW06 PW08 PW10 PW12 TW04 TW06 TW08 TW10 TW12

9.0 9.0 9.0 9.0 9.0 9.0 9.0 9.0

12 12 12 12 12 12 12 12 16 16 12 12 16 16 16 12 12 16 16 16

ο See Table 5 (right) for kW & amperage ratings ο Temperature control with an integral thermostat ο 316Lss wetted materials

NOTE: • It is strongly recommended when adding the Electric Heater to Maintain Temperature option to add the Disconnect with Start / Stop Buttons and Electric Heater Controller option, which adds the proper heater to pump safety interlocks. Without the controls package selected, it is the responsibility of the end-user to install the proper interlocks – failing to install and test an interlocking method between the pump and the heating element could lead to serious safety hazards and voids any warranty.

12.0 12.0

9.0 9.0

12.0 12.0 12.0

9.0 9.0

12.0 12.0 12.0

12

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER TECHNICAL DATASHEET

10.6 PUMP & MOTOR TYPE These options replace the recirculation pump’s standard TEFC motor with an upgraded rating. • Washdown Motor • Stainless Steel Washdown Motor

10.7 CONTROLS Adding a controls package to the base system allows for start/stop control of the pump and heater control if selected with the Electric Heater to Maintain Temperature option. An enclosure houses the components used to control the system and is equipment mounted. • The Disconnect with Start / Stop Buttons option package includes the following components: ο NEMA 4X rated control enclosure constructed of 304ss with a painted carbon steel back plate ο Two (2) Push buttons to start and stop the system

TABLE 6: DISCONNECT SIZE (AMPS)

460V AC No Electric Heat Electric Heat

Tank Series and Length

UW04 UW06 UW08 UW10 UW12 RW04 RW06 RW08 RW10 RW12 PW04 PW06 PW08 PW10 PW12 TW04 TW06 TW08 TW10 TW12

15 15 15 15 15 15 15 15 30 30 15 15 30 30 30 15 15 30 30 30

30 30 30 30 30 30 30 30 60 60 30 30 60 60 60 30 30 60 60 60

ο One (1) Motor starter/disconnect for the circulation pump ο One (1) Transformer to convert the incoming 3PH power to other needed voltages ο 20-ft. SO cord for the incoming electrical connection - a plug is not included ο Conduit (see Section 10.8 for options)

• The Disconnect with Start / Stop Buttons and Electric Heater Controller option package includes all the above components as well as: ο One (1) Fused disconnect switch (see Table 6 [right] for disconnect sizes) ο One (1) Heater controller ο One (1) Heater thermocouple amplifier for high temperature shutoff ο One (1) Selector switch for auto / off heater control NOTE: • Both options require selection of a conduit option (see Section 10.8 ). •  Disconnect with Start / Stop Buttons and Electric Heater Controller option requires selection of Electric Heater to Maintain Temperature option. • Both options add length (for control enclosure) to the system dimensions.

13

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER

10.8 CONDUIT The standard system has no controls associated with it and therefore no conduit type. When a controls package is selected, a selection of conduit type is then made. The For Electric Heater options are used when conduit is selected in addition to the Electric Heater to Maintain Temperature. • PVC Conduit or PVC Conduit – For Electric Heater : ο Rigid conduit: Schedule 40 nonmetallic PVC ο Flexible conduit and fittings: UL listed nonmetallic PVC with nylon fittings • Stainless Steel Conduit or Stainless Steel Conduit – For Electric Heater : ο Rigid conduit: Schedule 40 stainless steel ο Flexible conduit and fittings: UL listed PVC coated steel with stainless steel fittings With either conduit selection, instruments requiring calibration will include a minimum of 3 ft. of cable or flexible conduit allowing the device to be moved to a calibration cart without being unwired. NOTE: • All conduit options require selection of a controls package (see Section 10.7 ). •  PVC Conduit – For Electric Heater and Stainless Steel Conduit – For Electric Heater options require selection of the Electric Heater to Maintain Temperature option (see Section 10.5 ). 10.9 CHEMICAL MONITORING Selection of the Conductivity Analyzer option provides the opportunity for implementation of varying levels of chemical monitoring for the COP Parts Washer. The analyzer displays the chemical concentration of the wash solutions either by conductivity or percent concentration. It includes outputs that can be used by a chart recorder or a chemical dosing pump. When used with a chemical dosing pump, the chemical concentration is programmable energizing the pump when

additional chemical is needed (wiring by others). The included components are: • One (1) Rosemount 1056 series conductivity analyzer mounted to the equipment. • One (1) Rosemount 225 series conductivity sensor mounted in the sidewall of the tank.

10.10 REPORTING Chart recorders utilize circular paper charts to record certain cleaning parameters of the COP Parts Washer over a 24-hour period.

Two types of chart recorder options are available: • 1-Pen Chart Recorder, Temperature Recording ο One (1) Anderson AJ-300 1-Pen chart recorder for recording temperature is mounted to the equipment

ο One (1) ifm Efector TM-series RTD mounted in the sidewall of the tank • 2-Pen Chart Recorder, Temperature & Conductivity Recording ο One (1) Anderson AJ-300 2-Pen chart recorder for recording temperature and conductivity is mounted to the equipment ο One (1) ifm Efector TM-series RTD mounted in the sidewall of the tank NOTE: • The 2-Pen Chart Recorder, Temperature & Conductivity Recording option requires addition of the Conductivity Analyzer option (see Section 10.9 ). • Both options add length (for chart recorder) to the system dimensions.

14

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER TECHNICAL DATASHEET

The Data & Analytics brand from Sani-Matic brings real-time digital data capture and analysis to your fingertips. Using SaniTrend® software products, producers can monitor quality, ensure compliance, and drive efficiency across their cleaning processes. Options include:

• The Digital Process Recorder (DPR) Charts is a standalone control panel paired with an online Software-as-a-Service (SaaS) that provides digital, automated, and secure data acquisition and storage of critical process parameters. ο  DPR Charts automatically captures up to eight (8) analog values (e.g., temperature, conductivity) and up to twelve (12) digital values (e.g., pump running signal) and transmits them to a secure online portal for storage, review, and commenting. ο At the DPR HMI, users can see live data, review trending charts, and add comments to the online portal ο The analog and digital data can be reviewed in an easy-to-review chart for any time period, and users can add secure comments to a defined segment of data. ο The DPR connects and transmits data to the SaniTrend® Cloud online portal and can be accessed by your team securely through web-based platforms via desktop, mobile, or tablets – with no limit to the assigned users that can be created. ο The DPR Control Panel features include:

Powered by SaniTrend ® Cloud

» Large 7” HMI screen with clear data layouts » Hygienic Design Slope Top Control Panel » Trending Charted Data » Ability to add Comments to SaniTrend Cloud dataset locally at HMI

DPR Control Panel

» Networking configuration available at HMI » Local web server built in for HMI access

ο Additional features include live view dashboards, email and SMS (text) system alerts, and documentation storage all provide additional information and actionable insights into your process.

DPR Control Panel HMI Screen

SaniTrend® Cloud Online Portal View - Data Charts (DPR Charts)

NOTE: • This option also adds one (1) ifm Efector TD series temperature transmitter mounted in the sidewall of the tank, for collecting temperature data from the COP Parts Washer. • This option requires addition of the Conductivity Analyzer option (see Section 10.9 ) if conductivity recording is required.

• This option adds length (for the DPR control panel) to the system dimensions. • Find more information and tier comparisons on our SaniTrend® Cloud website page.

15

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER

10.11 FILL PORTS Selection of the Water and Chemical Fill Port option adds one (1) tri-clamp water fill port and one (1) 0.25” MNPT chemical fill port in the upper sidewall of the tank body. The fill port size is 1.5” TC for tanks with nominal volume ≤ 200 gallons, and 2.0” TC for nominal tank volumes > 200 gallons. NOTE: • Backflow prevention for the water fill port is by others (if required by the water supply).

11 ACCESSORIES

11.1 CHEMICAL PUMP ASSEMBLY, WILDEN P1 The chemical delivery system doses the cleaning chemicals into the tank. If the Conductivity Analyzer option is selected, the analyzer’s output can be tied into the air solenoid valve for dosing and maintaining a chemical concentration. For manual control, the air solenoid valve would be wired to an on/off switch. The below listed components of the assembly are loose shipped for installation by others. • One (1) PVC Suction lance (43”) with foot valve for placement into a 55-gallon drum (drum by others). • One (1) Wilden P1 series air operated diaphragm pump ο Duty: 4 gpm @ 65 psi

ο Instrument air requirement: 6 SCFM @ 80 psi ο Polypropylene body with PTFE diaphragms ο One (1) 0.25” Solenoid valve (120V AC) with filter / regulator for control of the pump. The pneumatic components and piping are made of brass. ο One (1) 304ss wall mounting bracket allowing the pump assembly to be placed near the chemical drum • One (1) Manual bleed valve to aid in priming the pump

Bleed Valve

Check Valve

• One (1) 0.25” Check valve to prevent the siphoning of the cleaning chemicals • One (1) 30 ft., 0.5” Polyethylene tubing for the pump’s suction and discharge lines • One (1) 8 ft., 0.375” Polyethylene tubing for the pump’s bypass/priming line NOTE: • It is recommended to add the Water and Chemical Fill Port option when using the chemical pump for a tie-in point for chemical injection.

Suction Lance with Foot Valve

16

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER TECHNICAL DATASHEET

11.2 END MOUNTED DRAIN TABLE The drain table provides a location for the washed components to dry or be reassembled. It has upturned lips to contain residual wash solutions and a downturn lip to drain back to the tank. The drain table is removable and can be placed anywhere along the perimeter of the tank. The base plate is 22" wide by 23" long. It is made of 304ss with weld color removed. NOTE: • This accessory cannot be used with the Hinged Cover option when the cover is closed (see Section 10.1 ). • This accessory must be mounted on the front or back of UW model tanks to accommodate its width.

11.3 TURN OVER PACKAGE (TOP) - HARD COPY A hard copy of the equipment Turn Over Package (TOP) will be printed and provided in a series of books (binders) . All Documentation materials listed in Section 9 are included in the hard copy package. 11.4 WIFI CONNECTION PACKAGE Adding the WiFi Connection Package enables connectivity for the DPR over local WiFi in place of Ethernet. This package includes an external, washdown rated WiFi hotspot installed on the DPR control panel. The following are included: • One (1) WiFi hotspot ο Washdown rated ο Mounted external to the DPR control panel ο Anybus model Wireless Bolt

17

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER

11.5 COP PARTS BASKETS

TABLE 7: COP PARTS BASKETS

Basket Dimensions (inches)

Handle Height (inches)

COP Without Cover COP With Cover UWRW PW TW UW RW PW TW

Model

Description

Part #

Placement: Tank bottom, non-stacking Construction: 316L sheet, 3 / 8 " diameter perforations, bead blast finish Placement: Tank bottom, non-stacking Common Uses: Fittings Construction: 316L round bar, 1" spacing, electropolished finish Placement: Edge of tank (hanger style) Common Uses: Small parts Construction: 316L sheet, 3 / 16 " diameter perfo- rations, electropolished finish Placement: Edge of tank (hanger style) Common Uses: Small parts Construction: 316L sheet, 3 / 16 ” diameter perforations, electropolished finish Placement: Tank bottom, non-stacking Common Uses: Small parts Construction: 316L round bar, 1" spacing electropolished finish Placement: Tank bottom, non-stacking Common Uses: Small parts Construction: 316L round bar, 1" spacing electropolished finish Placement: Tank bottom, non-stacking Common Uses: Small parts Construction: 316L round bar, 1" spacing electropolished finish Placement: Tank bottom, non-stacking Common Uses: Large parts Construction: 316L round bar, 1" spacing, electropolished finish Placement: Tank bottom, non-stacking Common Uses: Large parts Construction: 316L round bar, 1" spacing, electropolished finish Placement: Tank bottom, non-stacking Common Uses: Small parts. Fits in 5 gallon bucket for post-COP rinsing or sanitizing. Construction: 316Lss round bar, 0.22” spacing, electropolished finish

19.00 x 15.00 x 3.50

460030

ü ü ü ü ü ü ü ü

URWB-1

17.50

19.75 x 15.00 x 4.00

020132

ü ü ü ü ü ü ü ü

URWB-2

15.75

URWB-5-17 (Standard)

12.00 x 6.00 x 10.00

020139

ü ü ü ü ü ü ü ü

17.63

12.00 x 6.00 x 10.00

221298

ü ü ü ü ü ü ü

URWB-5-19

19.00

URWB-6-24 (Standard)

020136 22.00 x 10.50 x 10.00

ü ü ü ü

ü

24.00

218495 22.00 x 10.50 x 10.00

ü ü ü ü ü ü ü ü

URWB-6-15

15.00

218496 22.00 x 10.50 x 10.00 440021 22.00 x 22.00 x 18.00

ü ü ü ü ü ü ü

URWB-6-19

19.00

URWB-7-29 (Standard)

ü ü ü

ü

29.00

221293 22.00 x 22.00 x 18.00

ü ü ü ü ü ü

URWB-7-18

18.00

9.50” DIAM x 12.50”

062136

ü ü ü ü ü ü ü ü

URWB-8-12

17.00

NOTE: • The second number listed in the model designates the basket’s handle height (e.g., URWB-7-18 has an 18" handle height).

URWB-1

URWB-2

URWB-5-17

URWB-6-24

URWB-7-29

URWB-8-12

18

CLEAN-OUT-OF-PLACE (COP) PARTS WASHER TECHNICAL DATASHEET

SANIMATIC.COM HELLO@SANIMATIC.COM 800-356-3300

2855 INNOVATION WAY SUN PRAIRIE, WI 53590

LOOKING FOR MORE INFORMATION? Scan the QR code to explore everything SaniTech has to offer.

TD-0001.6